EasyManua.ls Logo

Omron K3HB - Page 125

Omron K3HB
150 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
User Calibr
User Calibr
ation
Section 6 User calibration
6-4
6.2 User calibration operation
Connecting to the Calibrator
Connect the Calibrator (standard voltage generator or standard current
generator) to the input terminal for the input type to be calibrated.
Use a Calibrator with enough precision for the accuracy of the K3HB-
S.
Key operation procedure
Perform the operation according to the following procedure.
Moving to calibration
level
Parameter Setting Procedure
The parameter character is "c
m
ov".
The parameter can be changed when the SV display starts to
flash.
If the password is correct, the Unit moves to the calibration level.
If the password is incorrect, the Unit remains in the advanced-
function setting level and the next parameter is displayed.
A Move to the advanced-function setting level, press the M[MODE]
Key several times and display the parameter to move to the
calibration level.
B Press the S[SHIFT] Key to make the SV display flash.
C Use the U[UP] and S[SHIFT] Keys to set the password. The
password is "1201"(1201).
D Press the M[MODE] Key to write the password.
Move to the calibration level.
The set value is always 0 after moving
from character display to monitor status.
(Does not move to the calibration level.)
To next parameter
Calibration level
Set value initialization
Password doesn't match.
Password matches.
Change status
cmo
cmo
0
0
u
adj
Use the
S
[SHIFT]
and
U
[UP] Keys
to set the password.

Table of Contents

Other manuals for Omron K3HB

Related product manuals