User Calibration
User Calibration
6.2 User calibration operation
6-5
Operation in
calibration level
Parameter Operation Procedure
A Use the procedure outlined above to
move to the calibration level.
The aging timer is displayed.
The aging timer is a 30-minute
countdown timer that counts until 0 is
reached.
A calibration record mark will be
displayed if a user calibration history
exists.
B Perform aging until the aging timer
reaches 0. (If the calibrator needs more
than 30 minutes of aging, extend the
aging until the conditions are met.)
If the M[MODE] Key is pressed while
the aging timer is counting down, the
display skips to the calibration upper
limit parameter display.
C Press the M[MODE] Key to display the
parameter for the calibration upper
limit.
The parameter for the present input
type is displayed. Refer to the table
on the next page for the relationship
between input type and parameters.
The display will be as shown below if
"A" is not included in the calculation
and the calibration will be for input
"B".
D The calibrator applies a reference
signal that corresponds to the
calibration upper limit.
Refer to the table on the next page
for required reference signal values.
E Press the U[UP] Key.
The reference signal is read and "T"
starts flashing.
F Press the U[UP] Key again to
temporarily register the calibration
upper limit.
If the M[MODE] Key is pressed
instead of the U[UP] Key, the upper
limit is not registered and the display
changes to the calibration lower limit
parameter.
adj
30
Displays "L
u".
L
u
30
.
adj
Calibration
record mark
L
u
adj
0
L
u
M
a4020
7f00
L
u
baa20
7f00
L
u
U
4000
7ef2
"T" is flashing.
L
u
U
4000
7ef2
"T" is lit.
L
u