EasyManua.ls Logo

Optek CONTROL 200 - Wiring of Ph Sensors

Optek CONTROL 200
135 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
C200 HARDWARE
Instruction Manual C200 Part1 A1.doc
optek Danulat GmbH D-45356 Essen Germany
page 18
*
4.3.2 Wiring of pH sensors
The converter CONTROL 200 can be connected to either a standard pH input or differential pH input. The
dual input pH is designed for use with pH electrode systems that incorporate a solution earth (ground)
connection to provide low glass impedance and out-of-sample/broken cable warnings.
The wiring of pH sensor PF12 with ABB combination electrodes to a converter C200 as sensor A and/or B is
described in the following tables:
Wiring with solution earth (ground) for inline sensor diagnostic
Terminal Block B
Sensor B Sensor A Wire no. Color Function
B1 B9 1 Grey Temperature Compensator Common
B2 B10 2 White Temperature Compensator 3
rd
lead
B3 B11 3 Green Temperature Compensator
B4 B12 4 Blue Reference
B5 B13 Not used Not used
B6 B14 6 Green/Yellow Solution Earth (Ground)
B7 B15 7 Red Screen/Shield
B8 B16 8 Transparent Glass Electrode
Wiring without solution earth (ground)
Terminal Block B
Sensor B Sensor A Wire no. Color Function
B1 B9 1 Grey Temperature Compensator Common
B2 B10 2 White Temperature Compensator 3
rd
lead
B3 B11 3 Green Temperature Compensator
B4 B12 Not used Not used
B5 B13 Not used Not used
B6 B14 4 Blue Reference
B7 B15 7 Red Screen/Shield
B8 B16 8 Transparent Glass Electrode
For use with different sensors refer to corresponding instruction manual !
If sensor diagnostics are required, ensure that Diff. Input for the relevant sensors is
set to “YES” – see Instruction Manual Control 200, Part 2 - Software !
If sensor is equipped with a 2-wire temperature compensator, link terminals B9 and
B10 (and B1 and B2 if dual input analyzer) !

Table of Contents

Related product manuals