EasyManua.ls Logo

Optek CONTROL 200 - Automatic, Single- and Two-Point Calibration

Optek CONTROL 200
135 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
C200 SOFTWARE
Instruction Manual C200 Part2 A1.doc
optek Danulat GmbH D-45356 Essen Germany
page 36
4.1.4 Automatic, Single- and Two-Point Calibration (pH Only)
Sensor A: Buffer Method (pH probes only)
Select the type of automatic calibration required:
Auto 1-Pt Automatic, single-point calibration
Auto 2-Pt Automatic, two-point calibration
Calibrate Buffer (Single-Point Calibration) or Calibrate Buffer 1(Two-Point
Calibration)
Immerse Sensor A in the buffer solution.
Press the
DOWNSCROLL key to initiate calibration.
Note. To abort calibration, press the DOWNSCROLL key again at any time before
calibration is complete – see below.
The center display line shows the measured sensor output in millivolts.
As calibration proceeds, a progress indicator appears in the lower display line. When
the measured sensor output stabilizes, the lower display line shows ##### 100%
#####.
The display then changes for 2 seconds to show the temperature-corrected buffer
value in the upper display line, then advances automatically to the next frame.
Two-point calibration selected – continued on next page.
Single-point calibration selected – continued on next page.
Abort Calibration
Select Yes or No.
Yes selected – return to the main menu.
No selected – calibration continues.
#### 100% ####
175
mV
A: Immerse Buf.2
A: Calibration
A: Abort Cal.
-----
Yes
A: SENSOR CAL.
A: Immerse Buf.2
A: Immerse Buf.1
4.00
pH
Deg.C
25.0
A: Buffer Method
Auto
-----
1-Pt
A: Cal. Buffer 1
4.01
pH

Table of Contents

Related product manuals