C200 SOFTWARE
Instruction Manual C200 Part2 A1.doc
optek Danulat GmbH • D-45356 Essen • Germany
page 82
5.2.7.2 Configure Power Failure Recovery Mode
Power Failure Recovery Mode
When power to the analyzer is restored, Control Mode (Section 2.2.2) is set
automatically according to the Power Failure Recovery Mode selected in this
frame.
Select the required mode:
Auto – Control Mode is set to Auto irrespective of its setting prior to the
power failure.
Manual – Control Mode is set to Manual irrespective of its setting prior to the
power failure. Control Output (Section 2.2.2) is set to the level set in
the Default Output frame below.
Last – Control Mode and Control Output are set to the same state as that
set prior to the power failure.
Default Output
Set the default output required after Power Failure Recovery, between 0 and
100% in 0.1% increments.
Note. A setting of 0% represents no output.
Return to main menu.
See Section 5.2.8.
Power Recovery
----
Power Rec. Mode
Default Output
50.0
----
%
Last
Manual
Auto
CONFIG. CONTROL
Power Recovery
CONFIG. SECURITY