EasyManua.ls Logo

Optek CONTROL 200 - Configure Diagnostics

Optek CONTROL 200
135 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
C200 SOFTWARE
Instruction Manual C200 Part2 A1.doc
optek Danulat GmbH D-45356 Essen Germany
page 48
5.1.4 Configure Diagnostics
Note. The Configure Diagnostics function is applicable only if Diff. Input for Sensor A and/or Sensor B is set to
Yes – see Section 5.1.3.
Configure Sensor A
Sensor B configuration (dual input analyzers only) is identical to
Sensor A configuration.
pH Glass Check
Set to On to enable the impedance of the glass electrode circuit to be monitored
in order to provide low glass impedance and out-of-sample/broken cable
warnings.
Note. This frame is displayed only if A: Probe Type is set to pH and A: Electrode
is set to Glass – see Section 5.1.3.
Reference Electrode Check
Set to On to enable the condition of the reference electrode to be monitored in
order to provide:
– an indication of when sensor cleaning or replacement is required
– out-of-sample/broken cable warnings
Sensor B (dual input analyzers only) configuration is identical to
Sensor A configuration.
See Section 5.1.5.
Reference Alarm
Set the impedance value above which the reference alarm is activated.
Sensor B (dual input analyzers only) configuration is identical to
Sensor A configuration.
See Section 5.1.5.
Config. Sensor A
-----
A: pH Glass
-----
A: Ref. Checking
-----
CONFIG. DIAGS
-----
Off
On
Off
On
A: Ref. Alarm
50
kOhm
CONFIG. ALARMS
Config. Sensor A
CONFIG. ALARMS
Config. Sensor B
Config. Sensor B
Config. Sensor B
Config. Sensor A
On
Off

Table of Contents

Related product manuals