EasyManua.ls Logo

Optek CONTROL 200 - Checking the Temperature Input

Optek CONTROL 200
135 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
C200 SOFTWARE
Instruction Manual C200 Part2 A1.doc
optek Danulat GmbH D-45356 Essen Germany
page 104
7.2.3 Checking the Temperature Input
Check the analyzer responds to a temperature input.
Disconnect the Pt1000/3K Balco leads and connect a
suitable resistance box directly to the analyzer inputs – see
Section 6.2.2. Check the analyzer displays the correct
values as set on the resistance box – see Table.
Incorrect readings usually indicate an electrical calibration
problem. Re-calibrate the analyzer as detailed in Section
6.2.3.

Table of Contents

Related product manuals