Overview 14: Configuration
151
Selects the display filter frequency for 3/5-Lead
ECG cable.
(Monitor)
(ST Monitor)
(EMS)
Selects the display filter frequency for the
attached therapy cable or 3/5-Lead ECG cable.
,
Selects the electrode label format. AAMI: RA,
LA, LL, RL, V; IEC: R, L, F, N, C.
,
Selects if HR/Arrhythmia alarms are on or off at
start up.
,
Determines if the rhythm status is seen on the
display with the primary ECG wave in all clinical
modes (except AED).
,
Selects the default High Alarm Limit for the HR
derived from the ECG and the pulse derived
from SpO
2
.
Adult: 35-300, 120 bpm
Infant/Child: 35-300, 160 bpm
adjusted in increments of 5
Selects the default Low Alarm Limit for the HR
derived from the ECG and the pulse derived
from SpO
2
.
Adult: 30-295, 50 bpm
Infant/Child: 30-295, 80 bpm
adjusted in increments of 5
Selects the VTach heart rate limit. Adult: 95-150, 100 bpm
Infant/Child: 95-150, 120 bpm
adjusted in increments of 5
Selects the VTach run limit. Adult: 3-20, 5
Infant/Child: 3-20, 5
Selects the PVC limit. Adult: 1-99, 10
Infant/Child: 1-99, 5
Table 33 Heart Rate/ECG Settings (Continued)
Parameter Description Setting Choices User Setting
Table 34 NBP Settings
Parameter Description Setting Choices User Setting
Selects the NBP color. , , , , ,
,
Selects the measurements units. ,
Selects if NBP alarms are on or off at start up. ,
Selects the frequency for NBP measurement:
manual or automatic on a selected schedule.
, q1, q2.5, q5, q10,
q15, q30, q60, q120
Selects the alarm source. , ,