User Manual
E727T0005, valid for E-727
BRO, 2019-06-28
Physik Instrumente (PI) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany Page 198 / 240
Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.ws, www.pi.ws
Parameter
ID (hexa-
decimal)
Data
Type
Command
Level for
Write
Access
Item Type Parameter
Name
Description
0x0e000e00 INT 1 System Adv. Piezo
Control License
Parameters of the Advanced
Piezo Control option (p. 19). See
the separate E712T0007
Technical Note for more
information.
0x0e000f00 INT 3 System Adv. Piezo
Control License-
Valid Flag
Input
The value of the parameter
enables/disables the Reset input
on pin 2 of the MDR14 socket
(details see p. 227).
0x0E001D00 INT 1 System Trigger Input
Filter Enable
The value of the parameter
enables/disables a filter for the
digital trigger input as follows:
0 = OFF: Filter is disabled.
Processing of the trigger input is
done immediately.
1 = ON: Filter is enabled to
reduce the influence of noise at
the trigger input line. Default
setting.
0x0E001E00 INT 1 System Discon. Target
Man. In With
Stop
Determines the behaviour when
the analog input is used as a
control source. See „Using the
Analog Input“ (p. 85) for details.
Position
Plausibility
Enables oder disables plausibility
check of sensor position.
0x0e002000 INT 1 System Move To Last
Commanded
Position
Determines the target positions
to be set when switching on the
servo mode:
0 = The target positions are set to
the current positions of the axes
(default).
1= The target positions are set to
the last commanded target
positions for the axes. This can
cause an abrupt motion.
0x0e002400 INT 1 System Trigger Input
Filter Samples
0x0f000000 INT 1 System Power Up Read
ID-Chip
For details, see “ID-Chip Support /
Stage Replacement” (p. 131).