User Manual
E727T0005, valid for E-727
BRO, 2019-06-28
Physik Instrumente (PI) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany Page 197 / 240
Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.ws, www.pi.ws
Parameter
ID (hexa-
decimal)
Data
Type
Command
Level for
Write
Access
Item Type Parameter
Name
Description
0x0e000200 FLOAT 3 System Servo Update
Time
[s]
0x0e000301 INT 1 System Disable Error 10 This parameter can be used to
avoid that error code 10 is set
when axes are stopped with the
STP, #24 or HLT commands.
0 = OFF (Error code 10 is set.)
1 = ON (Error code 10 is not set.)
0x0e000400 INT 1 System DDL license
valid
0x0e000b00 INT 3 System Number of
input channels
0x0e000b01 INT 3 System Number of
output channels
0x0e000b02 INT 2 System Number of
system axes
0x0e000b03 INT 3 System Number of
sensor channels
0x0e000b04 INT 3 System Number of
piezo channels
0x0e000b05 INT 3 System Number of
trigger outputs
0x0e000b06 INT 3 System Number Of
PiezoWalk
Channels
Not relevant for E-727.
0x0e000b07 INT 3 System Number of conf.
PiezoWalk chan.
Not relevant for E-727.
Filter
recommended with strain gauge
sensors and/or with servo
sampling rates ≤ 10 kHz.
Recommended parameter setting
for processing of strain gauge
sensors:
Cutoff frequency of the filter ≤
Servo sampling rate / 2
See response to HPV? for possible
values.