Planmeca ProMax 3D Max with ProTouch 57
TECHNICAL SPECIFICATIONS
User’s Manual
SmartPan 3D s / 3D Classic: 1.4 - 1.8
3D Plus / 3D Mid: 1.48 - 1.9
•Ceph 1.13
Duty cycle for height adjustment
25 s ON / 300 s OFF
Line voltage
100 - 220 V~ / 50 - 60 Hz
230 - 240 V~ / 50 Hz
Line current
8 - 17 A
Line harmonics
Cos better than 0.9
Max. permissible apparent
impedance of supply mains
0.5 (100VAC)
Maximum continuous heat
dissipation
< 250W
Fuses:
2 user replaceable fuses
100 - 220 V~ / 16A FF H 500V
230 - 240 V~ / 8A FF H 500V
Type
195100 ELU
Weight:
2D / 3D s / 3D Classic 119 kg (263 lbs)
3D Plus / 3D Mid 136 kg (300 lbs)
3D Max 134 kg (296 lbs)
Scanning ceph 26 kg (57 lbs)
Planmeca ProCeph 20 kg (44 lbs)
Colour
RAL 9016
Environmental requirements
Transport:
Temperature -20°C - +60°C
Relative humidity 10 - 90% RH (non-condensing)
Air pressure
700 - 1060 hPa
Storage:
Temperature -10°C - +50°C
Relative humidity 10 - 90% RH (non-condensing)
Air pressure 700 - 1060 hPa
Operating: