Powered by Safety
®
59
Appendix D
01.4IB.48041
Table AB MODBUS Memory Map (cont.)
Register Name
Read /
Write
Register
Type
Description Default
Range,
format:
UOM Register #
Modbus
Oset #
Interpretation
BriteSpot® Temperatures - 30330 to 30347
Temp C 1 R Input Temperature in C of Ch 1 N/A
-50 - 200,
-999,
0x7FFF
Temp 30330 331
(-)999 - Temperature invalid
Units depend on register 30027
0x7FFF - BriteSpot® module error
Temp C 2 R Input Temperature in C of Ch 2 N/A
-50 - 200,
-999,
0x7FFF
Temp 30331 332
Temp C 3 R Input Temperature in C of Ch 3 N/A
-50 - 200,
-999,
0x7FFF
Temp 30332 333
Temp C 4 R Input Temperature in C of Ch 4 N/A
-50 - 200,
-999,
0x7FFF
Temp 30333 334
Temp C 5 R Input Temperature in C of Ch 5 N/A
-50 - 200,
-999,
0x7FFF
Temp 30334 335
Temp C 6 R Input Temperature in C of Ch 6 N/A
-50 - 200,
-999,
0x7FFF
Temp 30335 336
Temp C 7 R Input Temperature in C of Ch 7 N/A
-50 - 200,
-999,
0x7FFF
Temp 30336 337
Temp C 8 R Input Temperature in C of Ch 8 N/A
-50 - 200,
-999,
0x7FFF
Temp 30337 338
Temp C 9 R Input Temperature in C of Ch 9 N/A
-50 - 200,
-999,
0x7FFF
Temp 30338 339
Temp C 10 R Input Temperature in C of Ch 10 N/A
-50 - 200,
-999,
0x7FFF
Temp 30339 340
Temp C 11 R Input Temperature in C of Ch 11 N/A
-50 - 200,
-999,
0x7FFF
Temp 30340 341
Temp C 12 R Input Temperature in C of Ch 12 N/A
-50 - 200,
-999,
0x7FFF
Temp 30341 342
Temp C 13 R Input Temperature in C of Ch 13 N/A
-50 - 200,
-999,
0x7FFF
Temp 30342 343
Temp C 14 R Input Temperature in C of Ch 14 N/A
-50 - 200,
-999,
0x7FFF
Temp 30343 344
Temp C 15 R Input Temperature in C of Ch 15 N/A
-50 - 200,
-999,
0x7FFF
Temp 30344 345
Temp C 16 R Input Temperature in C of Ch 16 N/A
-50 - 200,
-999,
0x7FFF
Temp 30345 346
Temp C 17 R Input Temperature in C of Ch 17 N/A
-50 - 200,
-999,
0x7FFF
Temp 30346 347
Temp C 18 R Input Temperature in C of Ch 18 N/A
-50 - 200,
-999,
0x7FFF
Temp 30347 348