Fig. 38: Menu CTFS input settings
SI1 - Conductivity
- Temperature -
Flow switch:
Sensor enabled
Enables/disables the CTFS sensor.
Conductivity Indicates the measuring range which is
defined for CTFS sensor.
Temperature Indicates the temperature range (0…50°C) of
the CTFS sensor which is defined for CTFS
sensor.
Starting delay
time
Specifies the time during which the conduc‐
tivity or temperature value is not to be taken
into account.
Adjustment possible from 0 to 480 minute(s).
Activation delay
time
Specifies the time during which the value must
remain stable before it is taken into account.
Adjustment possible from 0 to 240 second(s).
Deactivation
delay time
Specifies the time during which the value must
remain stable before it is taken into account.
Fig. 39: Menu analog input settings
CTFS input settings
4…20mA input settings
General Menu
39