COMMUNICATIONS PROTOCOL Cubiscan 325 command set
CubiScan 325 Service Manual 76
Weight units
This command is used to set the weight units to either English (pounds) or metric (kilograms)
mode.
Zero
This command should be issued periodically to force the Cubiscan 325 to perform internal
compensations to adjust to changes in temperature and humidity. This command should only be
3 1 Neg. acknowledge Alpha (N) 4Eh
4 1 End of text Control (ETX) 03h
5 1 Carriage return Control (CR) 0Dh
6 1 Line feed Control (LF) 0Ah
Pos Len Description Type Range ASCII
Command format
1 1 Start of text Control (STX) 02h
2 1 Wgt. unit command Alpha (#) 23h
3 1 English or Metric Alpha (E/M) 45h or 4Dh
4 1 End of text Control (ETX) 03h
5 1 Carriage return Control (CR) 0Dh
6 1 Line feed Control (LF) 0Ah
Acknowledge format
1 1 Start of text Control (STX) 02h
2 1 Wgt. unit command Alpha (#) 23h
3 1 Acknowledge Alpha (A) 41h
4 1 End of text Control (ETX) 03h
5 1 Carriage return Control (CR) 0Dh
6 1 Line feed Control (LF) 0Ah
Negative acknowledge format
1 1 Start of text Control (STX) 02h
2 1 Wgt. unit command Alpha (#) 23h
3 1 Neg. acknowledge Alpha (N) 4Eh
4 1 End of text Control (ETX) 03h
5 1 Carriage return Control (CR) 0Dh
6 1 Line feed Control (LF) 0Ah
Pos Len Description Type Range ASCII