RED RAVEN OPERATION GUIDE
COPYRIGHT © 2016 RED.COM, INC 955-0127_V6.3, REV-E | 101
CALIBRATE SENSOR: MANUAL CAPTURE
The camera offers two (2) ways to calibrate the sensor: Manual Capture and Auto Capture. This section
describes how to calibrate the sensor with Manual Capture. For more information on Auto Capture, go to
“Calibrate Sensor: Auto Capture” on page 102.
When you start a Manual Capture, the camera calibrates the sensor at the current exposure and temperature
setting. You will need to re-calibrate the sensor after extreme changes in exposure and temperature. The
Manual Capture process takes several minutes.
To perform a Manual Capture, follow the instructions below:
1. Ensure that the camera reaches the temperature you want to use for your shoot.
2. Set the exposure to the exposure you want to use for your shoot.
3. Ensure that the sensor is as dark as possible.
‒ Preferably, take off the lens, and install the front body cap.
‒ If you cannot take off the lens, put a lens cap on the lens. (Not all lens caps completely block IR. Lens
caps are made to protect the lens, not necessarily to block light.)
‒ If possible, use a dark room.
‒ Just closing the aperture is NOT sufficient.
4. Go to Menu > Settings > Maintenance > Calibrate > Create > Manual.
5. Select how many calibration maps you want to create from the Number of captures drop-down menu. You
can create up to four (4) calibration maps at a time.
6. Select the exposure you want for each calibration map from the Capture drop-down menus.
7. Select Start Capture(s).
8. Select OK to continue.
The system initializes and then makes multiple passes through the following phases of the calibration:
‒ Capturing
‒ Analyzing
‒ Erasing
‒ Programming
9. When the Calibration Successful dialog displays, select OK to complete black shading.
The new calibration map is automatically selected.