3.3 SERVICE TOOLS
RIELE BERLIN Photometer 5010 / V4 3.3.2 19.09.2003
DESCRIPTION OF SERVICE TOOLS
S1 System initialization
With this function the default parameters will be set.
Start menu by [MODE] [LF] [1]:
Confirm with [ENTER].
Following settings are then typed over in the battery powered RAM. They remain
active, even after switching the instrument off/on, until they are specifically
changed. This action ends with a long beep and a print-out of the program
release.
Beeper ON
Printer internal ON
Printer external OFF
Real time clock ON
Boost mode up to 390 nm
Language english
Analog-digital converter correction 75
Temperature control: Correction values 3000
EDP OFF
EDP protocol STX/ETX/BCC
Baud rate 9600 bps
Resulting factors of methods based on standard cleared
Error list cleared
Laboratory name not affected
User name not affected
Patients data not affected
Bubble detector ON
S2 Free methods initialization
Attention! With this function all methods edited by the user are deleted without call-back. In the case of doubt
these methods should be checked before: Scroll through all methods with the method editor and note them
down for an optional later new programming.
Start menu by [MODE] [LF] [2]:
Confirm with [ENTER].
This action ends with a long beep.
SYSTEM INIT. S1
SERVICE TOOLS
FR.METH.INIT. S2
SERVICE TOOLS