3.3 SERVICE TOOLS
RIELE BERLIN Photometer 5010 / V4 3.3.3 19.09.2003
S3 Parameter initialization
Attention!
The extended memory for the parameters of all nonlinear methods is deleted by this function without
call-back and the corresponding user defined methods become useless.
Start menu by [MODE] [LF] [3]:
Confirm with [ENTER].
This action ends with a beep.
S4 Serial number
Display of the current serial number. A new number can be entered with max. five digits.
Start menu by [MODE] [LF] [4]:
Confirm with [ENTER].
Confirm the input with [ENTER].
S5 Boost mode
To improve the low light gain at small wavelengths, the measurement time is extended when operating below
a definite wavelength. This definite wavelength will be displayed for approx. 2 sec. when starting this service
program. The default value is 390 nm.
Start menu by [MODE] [LF] [5]:
Confirm with [ENTER].
Enter [0] = BOOST OFF or a value between 300 and 900 as definite wavelength
in nanometers, below which the boost mode is activated. Finally confirm the input
with [ENTER].
S6 Filter select
The assignment of filter positions can be made.
Start menu by [MODE] [LF] [6]:
Confirm with [ENTER].
Select filter position by numeric key. Confirm the input with [ENTER].
Enter new wavelength by numeric key. Confirm the input with [ENTER].
Finish the operation with [BREAK].
PARAM. INIT. S3
SERVICE TOOLS
SERIAL NO. S4
SERVICE TOOLS
BOOST ON/OFF S5
SERVICE TOOLS
FILTERSELECT S6
SERVICE TOOLS