3.3 SERVICE TOOLS
RIELE BERLIN Photometer 5010 / V4 3.3.4 19.09.2003
S7 Automatic Result
With this service program a continuous measuring mode based on calculation procedure 1 (endpoint with
factor) is possible. After the input of wavelength, factor and delay-time and after measuring blank the con-
tinuous measuring mode is started.
Start menu by [MODE] [LF] [7]:
Confirm with [ENTER].
This mode is active until [BREAK] is pressed (and held for a while).
S8 Temperature functions
A sub menu with eight items for temperature functions is called.
Start menu by [MODE] [LF] [8]:
Confirm with [ENTER].
Direct selection of desired service tool, i.e. input of concerning item no. and
confirming with [ENTER], or by scrolling through all temperature functions
with [+] ... [+].
Item no. Service tool Description
in chapter
1 Temperature regulation on / off S8.1
2 Temperature value of standard
cuvette adaptor
S8.2
3 Temperature value of
flowthrough cuvette adaptor
S8.3
4 Temperature calibration S8.4
5 Temperature required / actual S8.5
6 Temperature Analog-Digital-
Converter
S8.6
AUTO. RESULT S7
SERVICE TOOLS
TEMP. FUNCT. S8
SERVICE TOOLS