EasyManua.ls Logo

Riele Photometer 5010 - S8.1 Temperature Regulation On;Off; S8.2 Temperature Value of Standard Cuvette Adaptor; S8.3 Temperature Value of Flowthrough Cuvette Adaptor

Riele Photometer 5010
114 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3.3 SERVICE TOOLS
RIELE BERLIN Photometer 5010 / V4 3.3.5 19.09.2003
S8.1 Temperature regulation on / off
The temperature regulation can be switched on or switched off. In the display the current setting is shown.
Start menu by [1]:
Confirm with [ENTER].
Changing of the setting with the keys [1] = ON or [0] = OFF.
The displayed setting becomes active with [ENTER] and [BREAK].
Select temperature level by [1], [2] or [3] in case of setting ON.
S8.2 Temperature value of standard cuvette adaptor
Display of the current setting. The correction of the temperature regulation is possible by change of the
temperature setting.
Start menu by [2]:
Confirm with [ENTER].
Select temperature level by [1], [2] or [3].
Enter new temperature value and confirm with [ENTER]
or finish operation with [BREAK]. The window shows for example the value 13047
for the temperature level 37 °C.
S8.3 Temperature value of flowthrough cuvette adaptor
Display of the current setting. The correction of the temperature regulation is possible by change of the
temperature setting.
Start menu by [3]:
Confirm with [ENTER].
Select temperature level by [1], [2] or [3].
Enter new temperature value and confirm with [ENTER]
or finish operation with [BREAK]. The window shows for example the value 13047
for the temperature level 37 °C.
TEMP. ON/OFF
INPUT +/E/B
HEATING OFF
1/ON 0/OFF
E-TEMP. VALUE
INPUT +/E/B
HEATING ON
1/25 2/30 3/37
E-TEMP. VALUE
1/25 2/30 3/37
37C VALUE E
13047
D-TEMP. VALUE
INPUT +/E/B
D-TEMP. VALUE
1/25 2/30 3/37
37C VALUE D
13047