6.1 COMMUNICATION PROTOCOL
RIELE BERLIN Photometer 5010 / V4 6.2 25.08.2003
List of commands:
Command
Action Answer
A [xxxxx]
Set: Device number
[xxxxx] = value
A
Query: Device number A [xxxxx]
e.g.: A3501
B
Query: Software version B [version, date of version /
version M1 / version M2]
e.g.:
BV4.0a 03.07.03 D/2.0/1.0
D
Set: Dark level adjustment
E [x]
Query: Dark values of
standard cuvette adaptor E and
flow-through cuvette adaptor D
[x] = multiplexer level 0...7
E [dark values]
e.g.:
E 111 D 123
F
Set: Measuring zero F [value]
e.g.: F 5.2739
F0
Set: Measuring zero with multiplexer
level select
F [value]
e.g.: F 5.2740
G
Set: Result without delay G [Result]
e.g.: G 0.3596
H [x]
Set: Multiplexer level
H
Query: Multiplexer level H [0], …, H [7]
e.g.: H6
I
Query: ADC value I [ADC value]
e.g.: I 174066
J
Query: Structure of method memory J [string length of a method;
number of basis methods;
number of programmable
methods;
number of pre-programmed
methods;
total number of methods]
e.g.: J 36; 19;230; 1;250;
K [method no.]
Query: Method string K [method no.;
method status;
calculation procedure /
temperature level;
wavelength;
no. of standards;
factor / standard;
volume
reserved
method name
unit
time 0
time 1
limit low
limit high
adress of parameters]
e.g.:
K 1; 19; 1; 0; 0; 0; 0;0;
C/F; ; 0; 1; 0; 0;
0;