EasyManua.ls Logo

Riele Photometer 5010 - List of Commands (Continued)

Riele Photometer 5010
114 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6.1 COMMUNICATION PROTOCOL
RIELE BERLIN Photometer 5010 / V4 6.3 25.08.2003
Command Action
Answer
L [method no.;
method status;
calculation
procedure /
temperature
level;
wavelength;
no. of
standards;
factor /
standard;
volume
reserved
method name
unit
time 0
time 1
limit low
limit high
adress of
parameters]
Set: Method string
N Query: Show actual position x of filter
wheel
N [1], …, N [9]
e.g.: N1
N [1], …, N [9] Set: Turn to position of filter wheel
O [xxx] Set: Wash: xxx µl
P [xxx] Set: Pump with bubble detector In case of a detected bubble
the error code is C 45 instead
of C.
Q Query: Language Q1 or Q2
Q1 Set: English language
Q2 Set: German language
R Set: Reset
S Query: Printer ON / OFF S [0], …, S [3]
e.g.: S1
S1 Set: Printer internal ON
S2 Set: Printer external ON
S3 Set: Printer internal and external ON
S0 Set: All printers OFF
T Set: Clear all programmed methods
U Query: Temperature U [0], …, U [3]
e.g.: U0
U0 Set: Temperature control OFF
U1 Set: Temperature control = 25 °C
U2 Set: Temperature control = 30 °C
U3 Set: Temperature control = 37 °C