6.1 COMMUNICATION PROTOCOL
RIELE BERLIN Photometer 5010 / V4 6.4 25.08.2003
Command
Action Answer
V1 Query: Temperature value 25 °C and
standard cuvette adaptor
V [xxxxx]
e.g.: V11570
V1 [xxxxx] Set: Temperature value 25 °C and
standard cuvette adaptor
V2 Query: Temperature value 30 °C and
standard cuvette adaptor
V [xxxxx]
e.g.: V12880
V2 [xxxxx] Set: Temperature value 30 °C and
standard cuvette adaptor
V3 Query: Temperature value 37 °C and
standard cuvette adaptor
V [xxxxx]
e.g.: V14740
V3 [xxxxx] Set: Temperature value 37 °C and
standard cuvette adaptor
V4 Query: Temperature value 25 °C and
flow-through cuvette adaptor
V [xxxxx]
e.g.: V10015
V4 [xxxxx] Set: Temperature value 25 °C and flow-
through cuvette adaptor
V5 Query: Temperature value 30 °C and
flow-through cuvette adaptor
V [xxxxx]
e.g.: V11270
V5 [xxxxx] Set: Temperature value 30 °C and flow-
through cuvette adaptor
V6 Query: Temperature value 37 °C and
flow-through cuvette adaptor
V [xxxxx]
e.g.: V13045
V6 [xxxxx] Set: Temperature value 37 °C and flow-
through adaptor
W Query: Required value of temperature
control
W [xxxxx]
e.g.: W 0
X Query: Actual value of temperature
control
X [xxxxx]
e.g.: X14102
Y Query: Stability of temperature control Y0 or Y1
Z Set: Suspend remote control After a delay of 5 seconds
the system sends a protocol
„M0“ in distances of
4 seconds until an „ACK“ is
received by an external
device.
a [xxxx] Set: Calibration of temperature
b Query: Beep ON / OFF b0 or b1
b0 Set: Beep OFF
b1 Set: Beep ON
c Reset system status
d Query: ADC of temperature control d [xxxxx]
e.g.: d 940
e Query: Type of cuvette adaptor
e1 = Standard cuvette adaptor
e2 = Flow-through cuvette adaptor
e1 or e2
f [x] Query: Settings of filter positions [x]
[www] = value of wavelength
f [x] [www]
e.g.: f3 546
f [x] [www] Set: Value of wavelength of filter position
g Set: Acoustic signal
h not used