Command Reference - Base Software
R&S
®
CMW500
232User Manual 1173.9463.02 ─ 06
Example:
CONFigure:FDCorrection:ACTivate RF1C,
'mytable_in', RX
CONFigure:FDCorrection:ACTivate RF1C,
'mytable_out', TX
Different tables are activated for the input and output direction of
the connector RF1C.
Example:
CONFigure:FDCorrection:ACTivate RF1C, 'mytable',
RXTX, RF1
The table is activated for paths using RF1C and RX1 or TX1.
Usage: Event
Firmware/Software: V1.0.5.3
V2.1.25: parameter <RFConverter> added
CONFigure:FDCorrection:DEACtivate <Connector>[, <Direction>[, <RFConverter>]]
Deactivates any correction tables for a specific RF connector or a specific connector /
converter combination. For bidirectional connectors the tables can be deactivated for both
directions or for one direction.
Depending on the installed hardware and the active sub-instrument only a subset of the
listed values is allowed. The mapping of virtual connector names to physical connectors
also depends on the active sub-instrument. For details see Signal Path Settings.
Setting parameters:
<Connector> RF1C | RF1O | RF2C | RF3C | RF3O | RF4C | RFAO | RFAC |
RFBC
RF1C, RF2C, RF3C, RF4C, RF1O, RF3O:
RF 1 COM to RF 4 COM and RF 1/3 OUT front panel connectors
RFAC, RFBC, RFAO:
Virtual names for the RF COM and RF OUT connectors
<Direction> RXTX | RX | TX
Specifies the direction for which the tables shall be deactivated.
RX means input and TX means output. For a pure output connec-
tor RX is ignored.
RXTX: both directions (for output connector only output)
RX: input (not allowed for output connector)
TX: output
Default: RXTX
<RFConverter> RF1 | RF2 | RF3 | RF4
RX and TX module in the path (RFn = RXn, TXn)
Example:
CONFigure:FDCorrection:DEACtivate RF1C
Deactivates all correction tables for the connector RF1C.
Example:
CONFigure:FDCorrection:DEACtivate RF1C, RX
Deactivates all correction tables for the input direction of connector
RF1C.