Command Reference - Base Software
R&S
®
CMW500
230User Manual 1173.9463.02 ─ 06
<Count> Maximum number of entries to be listed. By default all entries from
<StartIndex> to the end of the table are listed.
Return values:
<ValuePairs> Table entry pairs. Each pair consists of a frequency [Hz] and an
attenuation [dB].
Example:
CONFigure:BASE:FDCorrection:CTABle:DETails?
'mytable', 3, 1
Returns the entry with index number 3, e.g.: 2000000000,0.7
Usage: Query only
Firmware/Software: V1.0.5.3
CONFigure:BASE:FDCorrection:CTABle:LENGth? <TableName>
Returns the number of entries (i.e. pairs of frequency and attenuation) of a correction
table.
Query parameters:
<TableName> String parameter identifying the table. To display a list of existing
tables use the command CONFigure:BASE:FDCorrection:
CTABle:CATalog? on page 229.
Return values:
<TableLength> Number of table entries
Example:
CONFigure:BASE:FDCorrection:CTABle:LENGth?
'mytable'
For 'mytable' containing 4 frequency/attenuation pairs, the
result is 4.
Usage: Query only
Firmware/Software: V1.0.5.3
CONFigure:BASE:FDCorrection:SAV
Saves all correction tables of the addressed sub-instrument from the RAM to the hard
disk.
This action is performed automatically when the R&S CMW500 application software is
closed, e.g. by pressing the standby key. However, you can use the command to save
your work manually after creating, configuring or deleting correction tables.
Usage:
Event
Firmware/Software: V2.1.25
CONFigure:BASE:FDCorrection:RCL
Loads all correction tables of the addressed sub-instrument from the hard disk into the
RAM.