EasyManua.ls Logo

Rohde & Schwarz CMW500 - Trigger: B ASE: E XTA: D Irection; Trigger: B ASE: E XTA: S Ource; Trigger: B ASE: E XTB: D Irection; Trigger: B ASE: E XTB: S Ource

Rohde & Schwarz CMW500
323 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command Reference - Base Software
R&S
®
CMW500
246User Manual 1173.9463.02 ─ 06
Return values:
'<Source_1>' ...
'<Source_n>'
Comma separated list of all supported values. Each value is rep-
resented as a string.
Example:
TRIGger:BASE:EXTB:CATalog:SOURce?
Query the available output trigger signals for connector TRIG B.
A possible response is
"No Connection","Base1:
External TRIG A","Base1: External TRIG B".
Usage: Query only
Firmware/Software: V1.0.4.11
TRIGger:BASE:EXTA:DIRection <Direction>
TRIGger:BASE:EXTB:DIRection <Direction>
Selects the TRIG A and TRIG B connectors as either input or output connectors.
Parameters:
<Direction> IN | OUT
IN: Input connector
OUT: Output connector
*RST: IN
Example:
TRIGger:BASE:EXTB:DIRection OUT
Use TRIG B as an output connector.
Firmware/Software: V1.0.4.11
TRIGger:BASE:EXTA:SOURce '<Source>'
TRIGger:BASE:EXTB:SOURce '<Source>'
Selects the output trigger signals to be routed to the TRIG A and TRIG B connectors. The
signals listed below are always available. Depending on the installed options additional
values may be available. A complete list of all supported values can be retrieved using
TRIGger:...:CATalog:SOURce?.
Parameters:
'<Source>' 'No Connection': No output trigger signal available
'Base1: External TRIG A': Input trigger signal from the TRIG A
connector
'Base1: External TRIG B': Input trigger signal from the TRIG B
connector
*RST: 'No Connection'
Example:
TRIGger:BASE:EXTB:SOURce "Base1: External TRIG
A"
Select the input signal from TRIG A as an output trigger signal for
connector TRIG B.
Firmware/Software: V1.0.4.11
Instrument-Control Commands

Table of Contents