Command Reference - Base Software
R&S
®
CMW500
233User Manual 1173.9463.02 ─ 06
Example:
CONFigure:FDCorrection:DEACtivate RF1C, RXTX,
RF1
Deactivates all correction tables for paths using RF1C and RX1 or
TX1.
Usage: Event
Firmware/Software: V1.0.5.3
V2.1.25: parameters <Direction> and <RFConverter> added
CONFigure:FDCorrection:USAGe? <Connector>[, <RFConverter>]
Lists the correction tables assigned to a specific RF connector or a specific connector /
converter combination.
Depending on the installed hardware and the active sub-instrument only a subset of the
query parameter values is allowed. The mapping of virtual connector names to physical
connectors also depends on the active sub-instrument. For details see Signal Path Set-
tings.
Query parameters:
<Connector> RF1C | RF1O | RF2C | RF3C | RF3O | RF4C | RFAO | RFAC |
RFBC
RF1C, RF2C, RF3C, RF4C, RF1O, RF3O:
RF 1 COM to RF 4 COM and RF 1/3 OUT front panel connectors
RFAC, RFBC, RFAO:
Virtual names for the RF COM and RF OUT connectors
<RFConverter> RF1 | RF2 | RF3 | RF4
RX and TX module in the path (RFn = RXn, TXn)
If the specified converter value is incompatible with the connector
or the results are ambiguous because this parameter is omitted,
NAV is returned.
Return values:
<RXTableName> String identifying the table assigned to the RX direction. If no table
is active, an empty string is returned. For pure output connectors
the RX string is empty.
<TXTableName> String identifying the table assigned to the TX direction. If no table
is active, an empty string is returned. For pure input connectors
the TX string is empty.
Example:
CONFigure:FDCorrection:USAGe? RF1O
Result: "","mytable"
Usage: Query only
Firmware/Software: V1.0.5.3
V2.1.25: parameter <RFConverter> added
Instrument-Control Commands