Command Reference - Base Software
R&S
®
CMW500
208User Manual 1173.9463.02 ─ 06
SYSTem:PRESet [<Application>]
SYSTem:RESet [<Application>]
A PRESet sets the parameters of the current sub-instrument to default values suitable
for local/manual interaction. A RESet sets them to default values suitable for remote
operation.
The commands can be applied to the complete sub-instrument or to a selected applica-
tion of the sub-instrument.
Parameters:
<Application> String specifying the application to be reset/preset
Examples: "GPRF Meas2", "WCDMA Gen1", "LTE Sig1"
Example:
SYSTem:PRESet "GPRF Meas2"
Force the GPRF measurement instance 2 to a preset state opti-
mized for manual operation.
Usage: Event
Firmware/Software: V1.0.5.3
V3.0.10: [<Application>] added
6.3.6 Tracing the Remote Control Interface
The commands in this section configure and control tracing of the remote interface and
events into a file and enable/disable the display of the SCPI remote trace.
Before you start tracing, configure all settings as desired. Modifying settings while tracing
is active may result in loss of already traced data. Useful exception: Selecting a new target
file while tracing is allowed. For start mode "EXPLicit" a restart of the instrument resets
the settings to the documented default values.
If you want to start tracing already during start of the instrument, configure all settings
(including start mode "AUTO"). Then restart your instrument. Tracing will be started
automatically during the restart, using the already configured settings.
If you use an XML file as trace file, ensure that tracing is stopped properly. If tracing is
aborted instead of stopped, e.g. by shutting down the instrument for stop mode
"EXPLicit", the XML file will be invalid, because some tags are not closed.
When the maximum file size is reached (except for stop mode "BUFFerfull") or if tracing
is started with an already existing trace file, a backup of the trace file is created and the
file itself is reset and overwritten. When the file is full for the second time or tracing is
started the next time, the first backup file is lost because it is overwritten by the next
backup. In order to prevent loss of data, set a sufficient file size, select an appropriate
stop mode and archive/copy completed trace files if you want to keep them.
TRACe:REMote:MODE:FILE<inst>:ENABle......................................................................209
TRACe:REMote:MODE:FILE<inst>:FILTer.......................................................................209
TRACe:REMote:MODE:FILE<inst>:FORMat.....................................................................210
TRACe:REMote:MODE:FILE<inst>:NAME........................................................................210
TRACe:REMote:MODE:FILE<inst>:SIZE..........................................................................210
TRACe:REMote:MODE:FILE<inst>:STARtmode...............................................................211