Command Reference - Base Software
R&S
®
CMW500
255User Manual 1173.9463.02 ─ 06
Query parameters:
<Mode> FSCorrection | UCORrection | CALibration | OGCal
Type of the calibration for which information is queried
FSCorrection: Correction performed in factory or service
UCORrection: Correction performed by the user
CALibration: Verification in the factory
OGCal: Verification by the service (outgoing calibration)
Return values:
<Date> Date of the calibration as string
<Time> Time of the calibration as string
Example:
CALibration:BASE:LATest:SPECific? OGC
Queries information about the latest outgoing calibration. Possible
result: "2009-09-28","14:10:20"
Usage: Query only
Firmware/Software: V2.0.10
CALibration:BASE:ACFile?
Query name and creation date of the currently active RF path correction file.
Return values:
<Name> Name of the file as string
<Date> Creation date as string
Example:
CALibration:BASE:ACFile?
Possible result: "[Factory Default]", "2012-09-28"
Usage: Query only
Firmware/Software: V3.0.12
6.3.13 Status Reporting System
The STATus subsystem controls the SCPI-defined status reporting structures. For a
description of the status registers see chapter 5.6, "Status Reporting System",
on page 166.
The command description is structured as follows:
● General STATus Commands................................................................................256
● STATus:OPERation (Elementary Commands).....................................................256
● STATus:OPERation (Extended Commands)........................................................264
● STATus:OPERation (Overall Evaluation)..............................................................271
● STATus:QUEStionable.........................................................................................275
Instrument-Control Commands