Command Reference - Base Software
R&S
®
CMW500
245User Manual 1173.9463.02 ─ 06
SYSTem:BASE:SSYNc:MODE <Mode>
Specifies the role of the instrument in a multi-CMW setup, concerning the system time
synchronization signal.
Parameters:
<Mode> LIST | GEN | STAN
LISTener: The instrument receives a time synchronization signal
at SYS SYNC IN
GENerator: The instrument provides a system synchronization
signal at the rear panel
STANdalone: The instrument uses its internal synchronization
signal
Example:
SYST:BASE:SSYN:MODE GEN
Use the R&S CMW500 to generate a system synchronization sig-
nal.
Firmware/Software: V2.1.10
6.3.10.5 Trigger Settings
The following commands configure the connectors TRIG A and TRIG B at the rear panel
of the instrument and allows to initiate the generation of a "User Initiated Trigger" signal.
The settings are also available in the "Misc > Trigger" section of the "Setup" dialog.
Contents of this chapter
The commands in this section belong to the R&S CMW500 base system; they are not
related to specific firmware applications.
Most of the R&S CMW500 measurement firmware applications provide their own, specific
trigger settings. For details refer to the documentation of the TRIGger... subsystems
in the different firmware application chapters.
TRIGger:BASE:EXTA:CATalog:SOURce?........................................................................245
TRIGger:BASE:EXTB:CATalog:SOURce?........................................................................245
TRIGger:BASE:EXTA:DIRection......................................................................................246
TRIGger:BASE:EXTB:DIRection......................................................................................246
TRIGger:BASE:EXTA:SOURce.......................................................................................246
TRIGger:BASE:EXTB:SOURce.......................................................................................246
TRIGger:BASE:EXTA:SLOPe..........................................................................................247
TRIGger:BASE:EXTB:SLOPe..........................................................................................247
TRIGger:BASE:UINitiated<n>:EXECute...........................................................................247
TRIGger:BASE:EXTA:CATalog:SOURce?
TRIGger:BASE:EXTB:CATalog:SOURce?
Lists all trigger source values that can be set using
TRIGger:BASE:EXTA|EXTB:SOURce. The returned values depend on the intalled
hardware options and firmware applications.
Instrument-Control Commands