Remote Control
R&S
®
CMW500
120User Manual 1173.9463.02 ─ 06
1.
Connect your R&S CMW500 to the controller or to the home/company network using
the LAN REMOTE connector at the rear panel.
2.
Switch on the R&S CMW500, wait until the startup procedure is complete and press
the SETUP key.
The "Setup" dialog opens.
3. To the left, select section "Remote". Note the VISA address string displayed for
HiSLIP and close the dialog.
In the following, we assume that the following string is displayed:
"TPCIP::10.121.11.192::hislip0::INSTR"
4. Start your test tool. Configure the connection to the instrument using the VISA
address string and an alias, e.g. "CMW".
5. Write a test script using the alias and run the script.
Hostname and IP address
The VISA address strings displayed in the "Setup" dialog contain the hostname of the
R&S CMW500 instead of the IP address. Use the address string type that is most con-
venient for you.
The following test script queries the identification string of the connected R&S CMW500
and returns the contents of the error queue:
*IDN?
SYSTem:ERRor?
On test script execution, the test tool generates the following result log:
: Opening new VISA channel: TCPIP::10.121.11.192::hislip0::INSTR
: Connection to TCPIP::10.121.11.192::hislip0::INSTR established!
: Session handle: 0
: VISA Resource-Identifier: TCPIP::10.121.11.192::hislip0::INSTR
: send_Query(0, "*IDN?")
: [-->TCPIP::10.121.11.192::hislip0::INSTR] *IDN?
: read_Answer(0, ..., False)
: [<--TCPIP::10.121.11.192::hislip0::INSTR] Rohde&Schwarz,CMW,
1201.0002k50/624376,1.0.0.0
: send_Query(0, "SYSTem:ERRor?")
: [-->TCPIP::10.121.11.192::hislip0::INSTR] SYSTem:ERRor?
: read_Answer(0, ..., False)
: [<--TCPIP::10.121.11.192::hislip0::INSTR] 0,"No error"
5.1.3 Switching between Manual and Remote Control
On power-up, the instrument is always in the manual operating state. It can be operated
via the Graphical User Interface (GUI) and the (soft-)front panel controls.
Remote Control Operation