Basic Instrument Functions
R&S
®
CMW500
105User Manual 1173.9463.02 ─ 06
Fig. 4-16: Setup - Fan Control
Speed
Available modes:
"Normal"
This is the default mode.
"Low (Laboratory)"
The temperature thresholds for fan speed control are higher than in
normal mode. Generally this results in higher temperatures and less
noise emission.
So this mode can be used to minimize the noise emission, for example
in a laboratory environment.
"High (Production Line)"
The temperature thresholds for fan speed control are lower than in nor-
mal mode. Generally this results in lower temperatures and more noise
emission.
So this mode can be used to maximize the cooling and thus the lifetime
of the instrument components. This makes sense in an environment
where the noise emission of the instrument is irrelevant, for example at
a production line.
Remote command:
CONFigure:BASE:FCONtrol
4.5 Print Dialog
The "Print" dialog prints the current R&S CMW500 window to a file, to be saved on par-
tition D: of the internal hard disk or a USB memory stick. Other elements besides the
R&S CMW500 window are not captured (e.g. the desktop of the operating system or a
running program).
To open the dialog press the PRINT key at the (soft-)front panel.
Print Dialog