Command Reference - Base Software
R&S
®
CMW500
247User Manual 1173.9463.02 ─ 06
TRIGger:BASE:EXTA:SLOPe <Slope>
TRIGger:BASE:EXTB:SLOPe <Slope>
Specifies whether the rising edge or the falling edge of the trigger pulse shall be generated
at the trigger event.
The setting applies to output trigger signals provided at the TRIG A (EXTA) or TRIG B
(EXTB) connector.
Parameters:
<Slope> REDGe | FEDGe
REDGe: Rising edge
FEDGe: Falling edge
*RST: REDG
Example:
TRIGger:BASE:EXTB:SLOPe FEDGe
Generate a falling edge for TRIG B.
Firmware/Software: V3.0.10
TRIGger:BASE:UINitiated<n>:EXECute
Initiates the generation of a "User Initiated Trigger" signal. The corresponding trigger
source string is "Base1: User Trigger <n>".
Suffix:
<n>
.
1..2
User Trigger 1 or User Trigger 2
Example:
TRIGger:BASE:UINitiated2:EXECute
Generates a trigger pulse for the trigger source "Base1: User
Trigger 2".
Usage: Event
Firmware/Software: V3.0.10
6.3.10.6 Message Monitoring
The following command configures settings related to message monitoring. The settings
are also available in the "Misc > Logging" section of the "Setup" dialog.
CONFigure:BASE:MMONitor:IPADdress<n> <FirstSegment>, <SecondSegment>,
<NodeID>, <LocalID>
Configures the IP address pool for logging of signaling messages via an external PC.
The pool contains three IP addresses of external logging PCs.
The first two octets can not be configured. For a setting command you can specify any
values within the allowed range - they are ignored. A query returns the active values
resulting from the subnet configuration, see CONFigure:BASE:IPSet:SNODe
on page 249.
Instrument-Control Commands