Command Reference - Base Software
R&S
®
CMW500
249User Manual 1173.9463.02 ─ 06
SENSe:BASE:IPSet:SNODe:NTYPe?
Queries the subnet node type of the R&S CMW500.
Return values:
<Type> 'CMW'
Usage: Query only
Firmware/Software: V3.0.10
CONFigure:
BASE:IPSet:SNODe <ID>, <Description>, <NetworkSegment>
Configures the internal IPv4 subnet of the instrument.
Changing the <ID> or the <NetworkSegment> initiates a reboot of the instrument.
Parameters:
<ID> N1 | N2 | ... | N253 | N254
Selects the node ID, used as third octet of the IPv4 address.
N1 to N254 results in the IP address x.x.1.x to x.x.254.x.
Assign different IDs to all subnet nodes.
*RST: N1
<Description> String for easy identification of the subnet node, for example in the
subnet monitor output
*RST: 'My CMW'
<NetworkSegment> A | B
Selects the network segment to be used for the subnet
A: address range 172.22.y.z
B: address range 172.18.y.z
*RST: A
Example:
CONFigure:BASE:IPSet:SNODe N5, 'CMW room 5', A
The address range of the subnet equals 172.22.y.z. The node
itself uses IP addresses 172.22.5.z. In the subnet monitor the node
is listed as 'CMW room 5'.
Firmware/Software: V3.0.10
SENSe:BASE:IPSet:SNODe:NSEGment?
Queries information about the selected network segment and the resulting subnet prop-
erties.
Return values:
<SelectedSegment> A | B
Selected network segment
*RST: A