Command Reference - Base Software
R&S
®
CMW500
250User Manual 1173.9463.02 ─ 06
<IPAddress> String indicating the first two IP address octets for the subnet
*RST: '172.22.x.x'
<SubnetMask> String indicating the used subnet mask (fixed value)
*RST: '255.255.0.0'
Usage: Query only
Firmware/Software: V3.0.10
CONFigure:BASE:IPSet:NWADapter<n> <SetSubnetConform>
Assigns a subnet conform IP address to a network adapter of the instrument, selected
via index <n> or returns information about this network adapter.
A query returns <NWAdapterName>, <SetSubnetConform>, <IPAddress>, <Status>.
Suffix:
<n>
.
1..5
Parameters:
<SetSubnetConform> To assign a subnet conform IP address, send 1 or ON. To try
again, send first 0 or OFF, then again 1 or ON.
A query returns whether the last sent value was 0 or 1.
*RST: *RST has no effect on the value
Return values:
<NWAdapterName> Name of the network adapter as a string, e.g. "LAN Remote" for
<n> = 1
If OFF is returned, the selected value <n> is currently not assigned
to a network adapter.
<IPAddress> String containing the IP address (to be) assigned, see <Status>
<Status> NADJust | ADJust
State indicating whether the returned IP address has been suc-
cessfully assigned to the network adapter (ADJust) or not (NAD-
Just)
Example:
CONFigure:BASE:IPSet:NWADapter1 ON
CONFigure:BASE:IPSet:NWADapter1?
Assign a subnet conform IP address to network adapter 1 and
query information about the adapter, e.g. "LAN Remote",1,
"172.22.1.3",ADJ.
Firmware/Software: V3.0.10
SYSTem:BASE:IPSet:SMONitor:REFResh
Initiates an update of the information provided by the subnet monitor.
Usage:
Event
Firmware/Software: V3.0.10