Command Reference - Base Software
R&S
®
CMW500
211User Manual 1173.9463.02 ─ 06
Parameters:
<FileSize> Recommended minimum value: 40000 bytes
Maximum value: 1000000000 bytes (1 GB)
Default value: 1000000000 bytes (1 GB)
Example:
TRACe:REMote:MODE:FILE1:SIZE 100000000
Set 100 MB as maximum size for the trace file of sub-instrument
1.
Firmware/Software: V2.0.10
TRACe:REMote:MODE:FILE<inst>:STARtmode <StartMode>
Specifies whether tracing shall be started automatically or manually.
Suffix:
<inst>
.
1..2
Selects the sub-instrument
Parameters:
<StartMode> AUTO | EXPLicit
AUTO: Start tracing automatically when the instrument is started
EXPLicit: Start tracing via the command TRACe:REMote:
MODE:FILE<inst>:ENABle
Default value: EXPLicit
Example:
TRACe:REMote:MODE:FILE1:STARtmode AUTO
Enable automatic start of tracing for the next (re-)start of the
instrument.
Firmware/Software: V2.0.10
TRACe:REMote:MODE:FILE<inst>:STOPmode <StopMode>
Specifies how / when tracing shall be stopped and the trace file be closed.
Suffix:
<inst>
.
1..2
Selects the sub-instrument
Parameters:
<StopMode> AUTO | EXPLicit | ERRor | BUFFerfull
AUTO: Stop tracing automatically when the instrument is shut
down
EXPLicit: Stop tracing via the command TRACe:REMote:
MODE:FILE<inst>:ENABle
ERRor: Stop tracing when an SCPI error occurs
BUFFerfull: Stop tracing when the maximum file size is reached
Default value: EXPLicit
Example:
TRACe:REMote:MODE:FILE1:STOPmode BUFFerfull
Stop tracing when the maximum file size is reached (sub-instru-
ment 1).
Instrument-Control Commands

Table of Contents