TRACe:REMote:MODE:FILE<inst>:ENABle <Enable>
Enable or disable tracing of the remote interface to a file for the specified sub-instrument.
Suffix:
<inst>
.
1..2
Selects the sub-instrument
Parameters:
<Enable> Boolean value (ON | OFF | 1 | 0)
Default value: OFF
Example:
TRACe:REMote:MODE:FILE1:ENABle ON
TRACe:REMote:MODE:FILE2:ENABle OFF
Enable tracing for sub-instrument 1 and disable it for sub-instru-
ment 2.
Firmware/Software: V2.0.10
TRACe:REMote:MODE:FILE<inst>:FILTer <Input>, <Output>, <Error>, <Trigger>,
<DeviceClear>, <StatusRegister>, <Connection>, <RemoteLocal>
Specifies a filter for tracing of the specified sub-instrument. The filter defines which mes-
sage types and events are traced into a file.
The default setting is ON,ON,ON,OFF,OFF,OFF,OFF,OFF. All parameters support ON
| OFF | 1 | 0.
Suffix:
<inst>
.
1..2
Selects the sub-instrument
Parameters:
<Input> Trace incoming commands
<Output> Trace outgoing responses to queries
<Error> Trace SCPI error messages
<Trigger> Trace trigger events
<DeviceClear> Trace device clear messages
<StatusRegister> Trace status register changes
<Connection> Trace remote connection changes
<RemoteLocal> Trace transitions between local and remote operation mode
Example:
TRACe:REMote:MODE:FILE1:FILTer
ON,ON,ON,OFF,OFF,OFF,OFF,ON
Trace incoming commands, outgoing responses to queries, error
messages and remote-local transitions for sub-instrument 1.