9.2.3 Resetting values of the status reporting system
Table 10 comprises the different commands and events causing the status
reporting system to be reset. None of the commands, except for *RST,
influences the functional device settings. In particular, DCL does not change
the device settings.
Table 10: Resetting device functions. (1: the next command-line clears
the output buffer, DCL: Device Clear, SDC Selected Device Clear)
Effect Power
On
DCL,
SDC
*RST STATus:PRESet *CLS
Clear STB, ESR yes - - - yes
Clear SRE, ESE yes - - - -
Clear PPE yes - - - -
Clear EVENt sections of
the registers
yes - - - yes
Clear ENABle section of
all OPERation and
QUEStionable registers,
Fill ENABle sections of
all other registers with
"1".
yes - - yes -
Fill PTRansition sections
with "1" , Clear
NTRansition sections
yes - - yes -
Clear error queue yes - - - yes
Clear output buffer yes yes 1 1 1
Clear command
processing and input
buffer
yes yes - - yes
9.3 Error Messages
The following list contains all error messages for errors occurring in the
instrument. The meaning of negative error codes is defined in SCPI (Standard
Commands for Programmable Instruments), positive error codes mark errors
specific for the instrument.
In the left column the table contains the error text which is entered in the
error/event queue. In the right column there is an explanation regarding the
respective error.
For some errors, a so-called device-dependent info is added to the error
message. It gives further information about the error source (eg. -222, "Data
out of range", frequency too high).
Table 11: "No Error" message
Error code from queue
query