[SENSe]
. :MSCan
. . :LIST
. . . :STOP <numeric_value>|MINimum|MAXimum
Sets the last memory-list item that is used for a memory scan. The first list
item is set with SENS:MSC:LIST:STAR. A start location that is larger than the
stop location is rejected with error -221(“Settings conflict”).
Parameters:
<numeric_value> integer number in the range [0,1023]
MINimum|MAXimum set lowest|highest possible stop location
Example:
SENSe:MSCan:LIST:STOP 120
[SENSe]
. :MSCan
. . :LIST
. . . :STOP? [MINimum|MAXimum]
Query of number of memory list items used for direct save
Parameter:
<numeric_value> query of current stop location
MINimum|MAXimum query of lowest|highest possible stop location
Result:
integer number in the range [0,1023]
Example:
SENSe:MSCan:LIST:STOP? -> 120
11.14.2 Sense Panorama Scan subsystem PSC
The PSCan system controls the panorama-scan function of the device,
provided the panorama scan has been activated by
SENSe:FREQuency:MODE PSCan. Each scan is started by
INITiate[:IMMediate].
d
[SENSe]
. :PSCan
. . :COUNt <numeric_value>|MINimum|MAXimum|INFinity
Sets the number of RF-panorama scans. This command is an alias of
SENS:SWE:COUN and also changes its setting. This value is mapped onto
the same variable as the frequency scan and pscan count variable, meaning
that a change affects all scan modes.