3-3
Local Operator Interface
TABLE 3-1. LOI Data Entry Keys and Functions
Data Entry Keys Function Performed
Shift • Moves the blinking cursor on the display one character to the right.
• Scrolls through available values
Increment • Increments the character over the cursor by one
• Steps through all the digits, letters, and symbols that are applicable to the present operation
• Scrolls through available values
Enter Stores the displayed value previously selected with the SHIFT and INCR. keys
Display Control Keys Function Performed
Flow Rate Displays the user-selected parameters for flow indication
Totalize Displays the present totalized output of the transmitter, and activates the Totalizer group of keys
The choices, Forward and Reverse totals or Net and Gross totals, are selected in Auxiliary Functions
Totalizer Keys Function Performed
Start/Stop Starts the totalizing display if it is stopped, and stops the display if it is running
Read/Reset Resets the net totalizing display to zero if it is stopped, and halts the display if the display is running
Transmitter Parameters
Keys
Function Performed
Tube Calibration Number Identifies the calibration number when using Rosemount flowtubes, or other manufacturers’ flowtubes calibrated at the
Rosemount factory
Tube Size Specifies the flowtube size and identifies the corresponding maximum flow (0.15- through 60-inch line sizes)
Units Specifies the desired units:
Gal/Min Liters/Min
ImpGal/Min CuMeter/Hr
Ft/Sec Meters/Sec
Special (user defined)
Auxiliary Functions Function
Operating Mode
Coil Pulse Mode
Flow rate Display
Totalizer Display
Signal Processing
Special Units
Aux. Output Control
Universal Auto Trim
Coil Current (Model 8712U only)
Transmitter Gain (Model 8712U only)
Flowtube Gain (Model 8712U only)
Low Flow Cutoff
Pulse Width
Analog Output Zero
Analog Output Test
Pulse Output Test
Transmitter Test
4–20 mA Output Trim
Auto Zero
Electronics Trim
Options
Normal or Filter
6or30Hz
Flow–% Span, Flow–Totalize, %Span–Totalize
Forward–Reverse or Net–Gross
On/Off
Input Base Unit-Conversion-Time Base
Reverse Flow/Zero Flow
In-process Flowtube Calibration
75,125,500mA
A, B, or C
50 to 199.99 Percent
0.04 ft/s to 1 ft/s
Pulse Width
4 mA Value
Analog Output Loop Test
Pulse Output Loop Test
Test the Transmitter
Adjustthe4–20mAOutput
Zero Flow Tube for 30 Hz Coil Drive Operation
Transmitter Calibration
Analog Output Range Sets the desired 20 mA point – must set the tube size first
Pulse Output Scaling Sets one pulse to a selectable number of volume units – must set the tube size first
Damping Sets response time (single pole time constant), in seconds, to a step change in flow rate
Transmitter Information Allows you to view and change useful information about the transmitter and flowtube