5200 Multi Purpose and 5200 Multi Purpose HF Gas Analysers
05230001A / Revision 9
51
6.6.4 Select the Range associated with a measurement
Use the following procedure at any time during sampling to change the Range (see
Section 6.6.1) associated with a gas measurement:
1. Press the soft key to display the Menu screen, use the and soft
keys to highlight the "Set up" menu option, then press the soft key. The Set
up screen will then be displayed (see Figure 17).
2. Use the and soft keys to highlight the "mA output" menu option, then
press the soft key: the mA configuration screen is then shown (see
Figure 32).
3. Use the and soft keys to highlight the required gas Range option,
then press the soft key: the mA range screen will then be displayed, as
shown in Figure 34.
Figure 34 - The mA range screen
4. To change the selected Range:
• Press the soft key, then use the and soft keys to highlight the
alternative Range option (Range 1 or Range 2).
Press the soft key: the mA Range screen will then be displayed again, with the
new Range shown.