5200 Multi Purpose and 5200 Multi Purpose HF Gas Analysers
05230001A / Revision 9
69
Fault symptom Recommended actions
The Fault LED is on. Check any current fault messages (see
Section 8.2), and carry out the recommended
actions (see Section 8.1).
If there are no applicable fault messages stored, or
if you cannot rectify the fault after you have carried
out the recommended actions:
• Switch off the Multi Purpose, then switch it on
again.
• If the fault persists, contact Servomex or your
local Servomex agent for assistance.
The software health indicator is not
moving on the display.
Carry out the recommended actions for the "The
Fault LED is on" symptom above.
" - - - - - " is displayed instead of a
sample measurement (or appears in
the data log).
This indicates a possible measurement error, or a
communications error between a transducer and
the Multi Purpose controller.
Check that the Multi Purpose is not being knocked,
moved, or subjected to high levels of vibration
during sample measurements.
If the Multi Purpose is not being knocked, moved
or subjected to vibration and the fault persists,
contact Servomex or your local Servomex agent
for assistance.
" " is displayed instead of
a sample measurement (or appears
in the data log).
This indicates that the sample gas measurement is
above the upper measurement limit of the Multi
Purpose.
If this is displayed or appears during calibration,
check that the calibration gas is as specified in
Section 2.3 and recalibrate the Multi Purpose.
If this is displayed or appears during sample
measurement, check that the sample gas target
range is within the specification of the Multi
Purpose (see Section 2.2).
General fault finding (Sheet 1 of 5)