EasyManua.ls Logo

SICK GM32 Series

SICK GM32 Series
80 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
50
8012707/ZVF9/V2-1/2019-04 | SICKOPERATING INSTRUCTIONS | GM32
Subject to change without notice
6 MAINTENANCE
6.5 Cleaning the window
Fig. 32: SR-unit window
1 Swivel the SR-unit out, see “Swiveling out and removing the SR-unit”, page 49.
2 Cleaning the window.
Use an optical cleaning cloth for cleaning.
The cleaning cloth can be moistened with demineralized water.
Do not use cleaning agents.
3 Close the SR-unit again.
4 For GMP probe: Set the lever on the purge air fixture to the open” position again.
6.6 Checking and replacing the desiccant cartridges
Fig. 33: Desiccant cartridges
1 Swivel the SR-unit out, see “Swiveling out and removing the SR-unit”, page 49.
2 Desiccant cartridge is light blue: Desiccant cartridge is dry.
Desiccant cartridge is white: Replace the desiccant cartridge.
3 Replacing the desiccant cartridge:
a) Unscrew the desiccant cartridge.
b) Screw in the new desiccant cartridge.
4 Close the SR-unit again.
5 For GMP probe: Set the lever on the purge air fixture to the open” position again.
Window
2 desiccant cartridges

Table of Contents

Related product manuals