EasyManua.ls Logo

SICK GM32 Series

SICK GM32 Series
80 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
55
8012707/ZVF9/V2-1/2019-04 | SICK OPERATING INSTRUCTIONS | GM32
Subject to change without notice
CLEARING MALFUNCTIONS 7
7.2 Error diagnosis Tables
7.2.1 Device not functioning
7.2.2 Measured values clearly incorrect
7.2.3 Sample gas penetrating
Possible causes Information
Power supply not connected.
Check power cable and connections.
Main power supply has failed.
Check power supply (e.g. socket, external disconnecting device).
Internal operating temperatures
not correct.
Check whether error messages exist.
Internal software not functioning.
Can only occur with complex internal malfunctions or after strong external
influences (e.g. strong electromagnetic interfering pulse).
Swítch the GM32 off and back on again after a few seconds.
Table 14: Error diagnosis - device not functioning
Possible errors Possible causes Remedy
Leakages (on GPP probe). ---
Have the GPP probe checked by
SICK Customer Service.
Sample gas penetrated the area
before the SR-unit.
Gas pressure in gas duct too
high.
Purge air unit has failed or is too
weak.
see “Sample gas penetrating”,
page 55.
For GPP probe: Temperature dropped
below dewpoint.
---
Check project planning.
Sample gas penetrates the purge air
compartment.
Gas pressure in gas duct too
high.
Purge air unit has failed or is too
weak.
see “Sample gas penetrating”,
page 55.
Probe gap alignment incorrect.
Sample gas flow must not
pressed into the purge air
openings.
Check installation.
Sample gas conditions do not match,
or no longer match, project planning.
Plant condition has changed Check sample gas conditions
(temperature, moisture,
concentration etc.).
GM32 not ready for operation. ---
Check start-up
Check status/error messages.
GM32 calibration is not correct. ---
Check:
Were correct test gases used?
Are setpoint values correct?
If this does not help: Carry out a
calibration (please contact SICK
Service).
Analyzer is contaminated. ---
Report to manufacturer's Technical
Service or trained skilled persons.
Table 15: Error diagnosis - incorrect measured values
NOTE: Sample gas in the analyzer can damage the analyzer.

Table of Contents

Related product manuals