Specifications overview 9410 series
Pollution degree 2
Altitude < 3000 m (9843 ft)
Location/mounting Not suitable for wet locations
For indoor use only
Must be permanently connected and fixed
Maximum operating temperature
Max temp 948M2DO6DI
Digital (2 out, 6 in)
948M2AO4AI
Analog (2 out, 4 in)
70 ºC (158 ºF) 0 – 4 0,1
70 ºC (158 ºF) 0 2
60 ºC (140 ºF) 1, 2 2 – 4
LEDs
Energy pulsing LEDs (visible and infrared) located on top of meter
Type Visible orange, infrared
Wavelength 600 – 609 nm (orange) 825 – 925 nm (infrared)
Maximum pulse rate < 25 Hz (based on line frequency)
EMC (electromagnetic compatibility)
Product standards IEC 62052-11
IEC 61326-1
Immunity to electrostatic discharge IEC 61000-4-2
Immunity to radiated fields IEC 61000-4-3
Immunity to fast transients IEC 61000-4-4
Immunity to surges IEC 61000-4-5
Immunity to conducted disturbances IEC 61000-4-6
Immunity to voltage dips and
interruptions
IEC 61000-4-11
Immunity to magnetic fields IEC 61000-4-8
Immunity to ring waves IEC 61000-4-12
Immunity to conducted disturbances, 2 -
150 kHz
CLC/TR 50579
Radiated and conducted emissions FCC part 15 Class B, EN55011, EN55022 Class B, ICES-003 Class B
Surge withstand capacity (SWC) ANSI C37.90.1
7EN05-0336-03 197