Triple bend
3 x 90° bends in two planes
L2: Min. 20 x pipe diameter
L1: 3 x pipe diameter
Valves and pumps
Valves
L2: Min. 10 x pipe diameter, fully open valve
L1: 3 x pipe diameter
Partially opened valves
L2: Min. 40 x pipe diameter, partially opened valves (or equal valves design)
Pumps
L2: Min. 40 x pipe diameter
L1: 3 x pipe diameter
Installing/Mounting
4.2 Sensor installation
SITRANS FUS380/FUE380