Parameter lists
A
The following tables show the various parameters available via PDM.
The parameters are accessible at three different levels:
● Read: Maintenance level
● Read/Write (R/W): Specialist level (write access is only permitted after entering the user
password; default password is 1000)
● Hardware locked (HW key): The parameter is only accessible when a HW key is mounted
Note
For type-approved and verified FUE380 flowmeter the settings are HW key protected and
therefore read only. This HW key is protected via the verification sealing. The verification
sealing can only be opened by the user with the acceptance of the local authorities.
A.1 Identification
Table A-1 Identification parameters (FW 2.04 and EDD 1.02.09-08)
Parameter Default value Value range Access
level
Description
1 Application identifier Identity R/W Customer application identification informa‐
tion (max 16 characters)
2 Application location Location R/W Customer application location information
(max 16 characters)
Device
5 Sensor size Product de‐
pendent
R/W Pipe diameter of the sensor. It is only a text
field for measurement system information.
The value has no influence on the measure‐
ment. Please see inner diameter at pipe
date in parameter 306
8 Totalizer unit m
3
See Table A-6
Totalizer units
(Page 108)
R/W Totalizer unit as text for volume. Change the
unit in parameter table if the device menu
"Unit guide (offline)" does not list the wanted
unit.
Note: The device display can show "m
3
" unit
only - other units cannot be shown on the
display, but will be used and shown online
via PDM
SITRANS FUS380/FUE380
Operating Instructions, 05/2018, A5E00730100-AE 97