Note
Hardware (HW) key protected parameters
For type-approved and verified FUE380 flowmeter the settings are HW key protected and can
be only read, but not changed via PDM. The HW key is located behind the display and is
thereby protected by the display sealing. The verification sealing may only be broken with the
acceptance of the authorities.
In the following it is described how the device is commissioned using SIMATC PDM.
The commissioning is divided into the following steps:
1. Installing and connecting the IrDA interface adapter (Page 53)
2. Installing the device driver (Page 54)
3. Adding the device to the network (Page 56)
4. Configuring the device (Page 57)
5. Optimizing the system (Page 58)
6. Checking the operation readiness (Page 63)
6.5.1 Installing and connecting the IrDA interface adapter
Note
IrDA driver installation
For installation of the IrDA driver, refer to the instructions delivered with the adapter. The
required operating system to install SIMATIC PDM and the IrDA driver is Windows 7.
1. Connect IrDA adapter to PC
2. Mount adapter on FUS080
Commissioning
6.5 Commissioning via PDM
SITRANS FUS380/FUE380
Operating Instructions, 05/2018, A5E00730100-AE 53