Example: DN100, Q
p
= 60m
3
/h and Q = 1.2 m
3
/h
Ε
f
= ± (2 + 0.02 Q
p
/Q) = ± (2 + 0.02 (60/1.2)) = ± (2 + 1) = ± 3%
10.2 FUS380 and FUE380 systems
Table 10-1 Technical data SITRANS FUS380 and FUE380
Description Specification
Enclosure design/mate‐
rial
Fiber glass reinforced polyamide in light-gray color
Enclosure IP67 according to EN 60529 and DIN 40050 (NEMA 4X/6)
Wall mounting kit Terminal box made of fiber glass reinforced polyamide in light-gray color with
cable glands and a stainless steel bracket for wall or pipe mounting
Cable glands In wall mounting kit:
3 x M20 PA plastic cable glands (1 for mains power supply cable, and 2 for
output cables)
Ambient temperature
MID version: -10 °C to +55 °C (14 °F to 131 °F)
Non-MID version: -10 °C to +60 °C (14 °F to 140 °F)
Storage temperature −40 °C to + 85 °C (−40 °F to +185 °F)
Sensor cables Coaxial cable, impedance 75 Ohm, ∅ app. 6 mm
Connection cable Cable length:
Max. 30 meter between transmitter and sensor terminal box
(5 m (16.4 ft), 10 m (32.8 ft), 20 m (65.6 ft), or 30 m (98.4 ft) available)
Mechanical vibration 2 g, 1 to 800 Hz sinusoidal in all directions according to IEC 68-2-6
Power supply Battery
1)
:
Replaceable 3.6 V LiSOCl (Lithium Thionyl Chloride)
Double battery pack 33 Ah or single back-up battery 16.5 Ah;
Battery change interval: 4.2 / 6 years at 60 °C (140 °F) operation
Mains:
87 to 265 V AC (50 to 60 Hz)
Display
LCD, 8 digits, additional 2 digits and symbols for status information
Units: Volume unit: m
3
(default display unit), Flow unit: m
3
/h (default
display unit)
Alarm codes (F 1 to F 9) for: path 1 or 2 not measuring, internal failure,
power supply failure, flow overload, pulse output frequency overload, data
logger warning
Push button One push button for toggling between display information
Measuring function
0.5 Hz battery mode
or
15 Hz mains-powered
Communication IrDA on display panel (Modbus RTU protocol)
Technical data
10.2 FUS380 and FUE380 systems
SITRANS FUS380/FUE380
88 Operating Instructions, 05/2018, A5E00730100-AE