1. Replace blind plug with cable gland (if two output signals are required, use a double entry
cable gland).
2. Push output cable(s) through open gland.
3. Connect output cable(s).
([WHUQDOFRQQHFWLRQ
&RQQHFWLRQ9DULDQW
3DVVLYHRXWSXW
2SHQGUDLQ



2XWSXW$
2XWSXW%
3RVLWLYHORJLF
5
5
9ื9'&
9




9
9
9
,P$
,P$
9ื9'&
6LJQDO
6LJQDO
4. Tighten cable gland for output cable(s) (approx. 20 Nm).
Note
Cables with voltage
Keep the output signal cable(s) separated from cables with voltages > 60 V.
Note
Ingress protection
To guarantee the IP67 (NEMA 4X/6) degree of protection, use cables with the required
specifications.
Note
Protected terminal
The wire size for the mains-powered terminals are AWG24 to AWG16 or 0.205 mm
2
to
1.500 mm
2
.
Note
EMC performance
Incorrect fixing of the output cable shield will affect the EMC performance.
Connecting
5.4 Remote system
SITRANS FUS380/FUE380
42 Operating Instructions, 05/2018, A5E00730100-AE