Sirona Dental Systems GmbH 4Technical data
Service Manual INTEGO / INTEGO pro
båÖäáëÜ
4
Technical data
Model designation, power ratings
Unit classifications, without HF surgery
Year of manufacture
Minimum PC requirements
Model designation: INTEGO / INTEGO pro
Power connection: 100 240 V AC ± 10%
50/60 Hz
Rated current: 3.3 A–1.5 A at 100–240 V
also max. 6 A for external devices
Type of ground connection: TN-C-S system or TN-S system
(acc. to IEC 60364-1)
Overvoltage category: 2 acc. to IEC 60664-1
Average power consumption
(for dimensioning an air
conditioning system):
100 W
Power consumption when
switched off:
0 W (power switch present)
Main building fuse: Type B automatic circuit breaker
100 - 115 VAC: 20 A slow-blow
220 240 VAC: 16 A slow blow
Protection class: Class Idevice
Device class in accordance
with Directive 93/42/EEC:
Class IIa equipment
Degree of protection against
electrical shock:
Type B applied parts
External intraoral camera SiroCam F /
AF / AF+. Namely:
Applied part type BF
Degree of protection against
ingress of water:
Ordinary equipment (without protection
against ingress of water)
The foot switch has an IPX1 degree of
protection against liquids (drip-proof).
Year of manufacture:
(on the rating plate)
USB port: corresponds to USB 2.0 standard
IMPORTANT
Minimum requirements for PC
See document "Installation instructions and system requirements for PC
configuration," (REF 61 94 075 SIVISION digital.